MANSON MODEL 5 MULTI-ANODE ULTRASOFT X-RAY SOURCE
The source is shown below viewed from the "load-inspection port"
side with the two exit beams directed away at 45 degrees through the two gate
valves. The anode selection is made by rotation of the handle at the top
right, and the two beam filters are controlled by the micrometer handles shown
at the top rear. The three ports shown on this side each are covered by a
pyrex window to allow inspection of anodes and the gearing mechanism under
vacuum. This allows a visual check of any of the anodes to see if any obvious
degradation has taken place, without having to break the vacuum, and also
check the moving parts in case some problem is suspected. Anode caps are
replaced by removal of the 2-3/4 inch viewport in the center, so the 8"
top flange seal is seldom opened.
The x-ray output and electrical characteristics of the source are the same
as for our Model 2 X-ray Source. It will operate at
up to 10 kV with a loading of up to 10 watts electron beam power. The source
spot is approximately 200 microns in diameter at 10 kV and 0.1 mA, increasing
to around 2 mm diameter at 5 kV and 2.0 mA. Spot diameter increases with the
beam current, and decreases with increasing beam voltage. Although normal
practice is to turn off the HV before rotating the 6-position anode carousel,
this selection of anodes may take place with the high voltage applied. Very
rapid cycling of the emitted radiation can therefore be accomplished. The
built-in beam filters are mounted on linear actuators to allow a position
close to the source spot, for a fully filtered beam of 0.1 ster- radian
aperture which fills the 2.5 inch ports of the exit beam gate valves, 7 inches
from the source spot.
Included with the basic x-ray source is the vacuum housing shown here, with
gate valves, filter actuators including 150 µg cm-2 VYNS and 6
micron Polypropylene filters for both exit beams. Spare filter holders (8) are
included, in addition to 4 spare filaments, two sets of 6 anode caps, each
comprised of Al, Mg, C, MgO on Mg, B on Mg, and Be/MgOH on Mg caps. Two
additional Al and two Mg caps are mounted for shipment, along with one C and
one MgO on Mg caps. One of the Al caps will have the typical electron beam
signature spot which has been deposited during testing, to provide the
customer with a reference for comparison during his familiarization tests of
the instrument. A 12 foot HV cable with 10 kV plug termination, to mate with
the 10 kV receptacle on the source, is also provided. The customer will attach
the plug which mates with his HV supply to the other end of this cable. A
regulated and filtered l0 KV supply such as the Bertan
Model 205B-1OR is recommended. The filament heating power is provided by
the Model 3B X-ray Source Control, included with
the Model 5, through the 12 foot filament current cable.
The x-ray output of the source at the
various light element K-bands and lines is the same as for our Model 2 source,
that is, up to 1012 photons s-1 sr-1.